• Log in / Register | My account | Log out | Downloads Safety data sheet PAK Marker

  • Water Cooling Unit – Interlab B.V.

    Water Cooling Unit

    can be installated inhouse or outside (not in direct sun exposition) providing cooling water for up to 2 analyzers, extraction machines etc. Closed water cooling system with tank and electronically controlled cooling water temperature with digital …..

    Description

    Water Cooling Unit can be installated inhouse or outside (not in direct sun exposition) providing cooling water for up to 2 analyzers, extraction machines etc. Closed water cooling system with tank and electronically controlled cooling water temperature with digital display. In case of installation outside at temperatures below 0°C the tank has to be filled with an anti-freezing water-/glyclol mix, which should be not less than 25%. The cooling capacity of the unit is sufficient for the operation of up to two analyzers at same time. A suitable tubing installation has to be provided locally. Layout details available on request.

    • Water tank capacity 26 l
    • Refrigeration capacity at ambient temperature of 32°C = 4000 resp. 6670 W at flow temperature of medium 0 resp. 10°C
    • admissible ambient temperature -20 .. +42°C
    • air capacity 4500 m³/h
    • flow rate of circulation pump 2 m³/h
    • flow height of circulation pump 3,1 bar

    Additional information

    Electrical data

    400 V, 50 Hz, 3 P+N+PE, 4.4 kW, 16 A

    Dimensions approx.

    720x735x930 mm

    Weight approx.

    130 kg

    Standards

    —–

  • Interlab BV | Europalaan 21 | NL-4645 HP Putte | Tel: +31 76 50 22 540 | Email: info@interlab-bv.nl | VAT: NL006965581B01 | CoC Number: 20047805